System Phantom Lite T1, T2, PD Values

≥129-0001 T1 Data

Measured officially at NIST Boulder at 3.0 T between 16°C and 26°C, and at 1.5 T at CMRI Headquarters at approximately 20°C.

Data is relevant to Premium System Phantom serial numbers ≥129-0001.

NIST Traceable

Data measured and published by NIST Boulder, the Magnetic Imaging Group (Report of Measurement: Magnetic Resonance Measurements for MRI Biomarkers: Proton Spin Relaxation Times, NIST Boulder, Certificate Number 80100S-20200610-C1).

Calibrations were for solutions and phantoms provided to the NIST/NIBIB lending library.

Click here to access this report of measurement.

Please wait while flipbook is loading. For more related info, FAQs and issues please refer to DearFlip WordPress Flipbook Plugin Help documentation.

129-0012 T2 Data

Measured officially at NIST Boulder at 3.0 T between 16°C and 26°C, and at 1.5 T at CMRI Headquarters at approximately 20°C.

Data is relevant to System Phantom Lite serial numbers 129-0012.

Data measured and published by NIST Boulder, the Magnetic Imaging Group (UNOFFICIAL REPORT of Measurement: Magnetic Resonance Measurements for MRI Biomarkers: Proton Spin Relaxation Times, NIST Boulder, Certificate Number 80100S-20200610-C1)

This report is marked unofficial because the T2-1 and T2-5 had to be remeasured because of supplier manufactur error. All data points besides the ones mentioned above are accurate and correct.

Click here to access this unofficial report of measurement.

Please wait while flipbook is loading. For more related info, FAQs and issues please refer to DearFlip WordPress Flipbook Plugin Help documentation.

≥129-0013 T2 Data

Measured officially at NIST Boulder at 3.0 T between 16°C and 26°C, and at 1.5 T at CMRI Headquarters at approximately 20°C.

Data is relevant to System Phantom Lite serial numbers ≥129-0013.

NIST Traceable

Data measured and published by NIST Boulder, the Magnetic Imaging Group (Report of Measurement: Magnetic Resonance Measurements for MRI Biomarkers: Proton Spin Relaxation Times, NIST Boulder, Certificate Number 80100S-20200610-C1).

Calibrations were for solutions and phantoms provided to the NIST/NIBIB lending library.

Click here to access this report of measurement.

Please wait while flipbook is loading. For more related info, FAQs and issues please refer to DearFlip WordPress Flipbook Plugin Help documentation.

≥129-0001 Proton Density Data

Measured officially at NIST Boulder at 3.0 T at approximately 20°C.

Data is relevant to Essential System Phantom serial numbers ≥129-0001.

NIST Traceable

Data measured and published by NIST Boulder, the Magnetic Imaging Group (Report of Measurement: Magnetic Resonance Measurements for MRI Biomarkers: Proton Spin Relaxation Times, NIST Boulder, Certificate Number 80100S-20200610-C1).

Calibrations were for solutions and phantoms provided to the NIST/NIBIB lending library.

Click here to access this report of measurement.

Please wait while flipbook is loading. For more related info, FAQs and issues please refer to DearFlip WordPress Flipbook Plugin Help documentation.

≥129-0001 Copper Sulfate Data

Measured officially at NIST Boulder at 3.0 T between 16°C and 26°C.

Data is relevant to Essential System Phantom serial numbers ≥129-0001.

NIST Traceable

Data measured and published by NIST Boulder, the Magnetic Imaging Group (Report of Measurement: Magnetic Resonance Measurements for MRI Biomarkers: Proton Spin Relaxation Times, NIST Boulder, Certificate Number 80100S-20200610-C1).

Calibrations were for solutions and phantoms provided to the NIST/NIBIB lending library.

Click here to access this report of measurement.

Please wait while flipbook is loading. For more related info, FAQs and issues please refer to DearFlip WordPress Flipbook Plugin Help documentation.